As ultrasound we describe acoustic waves with frequencies higher than 20 kHz, which propagate through the material environment on the principle of particle densification and dilution. During its transmission, no electric current passes, so we refer to its therapeutic use as mechanotherapy.

Principles of the therapeutic effect of ultrasound[edit | edit source]

The therapeutic effect of ultrasound is due to the fact that it is absorbed to a different extent by different substances.

  • Absorption coefficient is then directly proportional to the effectiveness of the treatment. It increases with increasing protein content in the tissue, and decreases with increasing water content. The absorption coefficient of a frequency of 1 MHz is three times smaller than the absorption coefficient of a wave of a frequency of 3 MHz and penetrates three times deeper. Therefore, ultrasound with a frequency of 3 MHz is used for more superficially located tissues, while ultrasound with a frequency of 1 MHz, on the other hand, has greater effects with deeper structures, and has almost no effect on the surface.
  • At the tissue interface, ultrasound waves are reflected to varying degrees - the reason is the different specific impedances of the individual tissues. Air has a very high impedance, its 10 nm thick layer reflects up to 99% of ultrasound waves.[1]
  • The Effective radiating area (ERA) of the head' is determined by the size of the piezoelectric crystal generating the ultrasound. As a rule, it reaches values ​​of 1−10 cm2. Based on the distance from the warhead, we distinguish an ultrasound near field characterized by low divergence of the ultrasound "beam" and relatively large intensity variations due to strong [[w:cs:interference|interference effects] ] (i.e. strengthening or weakening of ultrasound energy). On the contrary, distant ultrasound field is characterized by high beam divergence and lower intensity.
  • Half-depth of penetration is the distance at which the ultrasound intensity drops to 50% of its original size.[1]
  • Penetration depth is the maximum depth at which a therapeutic effect can be expected (the wave intensity in this place is ten times smaller than the wave intensity close to the head).[1]

Usage examples[edit | edit source]

Use in stomatology[edit | edit source]

Magnetostrictive sources are used as the source of ultrasound waves, which distinguishes dentistry from other fields where ultrasound is used therapeutically. The source of the oscillations here is a ferromagnetic rod, which changes its volume in rapid succession due to the action of the magnetic field. The advantage of this method of generation is high performance, however, this can only be achieved with frequencies up to 60 kHz'[2]. Frequencies in the range of 24-42 kHz (low-frequency ultrasound) with an intensity of more than 10 W/cm2[3], to remove tartar and cavity. Compared to manual removal, this method is disproportionately faster, less demanding and saves enamel. The main factor that destroys tartar is the formation of cavitation in the water that runs down the tool itself, which is made of titanium.[3]

Hyperthermia[edit | edit source]

Ultrasound is, in addition to microwave radiation, one of the ways to achieve overheating of the tumor and its subsequent damage (see hyperthermia for more). The tumor must be exposed to a temperature of 41-43 °C for about 20 minutes [3]. The targeted effect is achieved by focusing, i.e. by concentrating ultrasound waves coming from different directions to the tumor site; therefore, there is minimal damage to healthy tissue. A frequency of 1 MHz is usually used, the temperature is always controlled thermocouples.

Harmonic Scalpel[edit | edit source]

Harmonic scalpel works on the principle of longitudinal ultrasonic oscillation at a frequency of 55 MHz. This creates energy at low temperatures from 50 to 100 °C. It allows the cutting and coagulation of soft tissue, blood vessels and sealing of lymphatic channels in one moment. Bleeding is 4x to 6x less when using the harmonic scalpel than with other surgical methods. This is a very gentle surgical technique used especially in plastic and endoscopic surgery, near nerves and blood vessels.

Physical therapy[edit | edit source]

One type of physical therapy that is very widespread is ultrasonotherapy, which uses the acoustic energy of ultrasound. Ultrasound is applied using a special device equipped with a radiation head with an effective surface of various sizes from 1 to 10 cm2. It forms mechanical waves with a therapeutic frequency of 0.7 to 3.3 MHz, which is maximally absorbed at a tissue depth of 2 to 5 cm. Therefore, it is used in the therapy of soft tissue damage. It has an analgesic, spasmolytic and vasodilating effect.

Application[edit | edit source]

We divide according to the movement of the head:

  • static (the head is applied to one affected place and does not move with it);
  • semistatic (circular movement around the affected area);
  • dynamic (movement of the head, e.g. along the entire limb).

Next, we divide the heads according to contact with the skin:

  • direct contact of the head with the skin (using gel or oil);
  • contact through the aquatic environment.

The application usually takes a few minutes and is repeated several times over the course of consecutive days.

Indication[edit | edit source]

Contraindications[edit | edit source]

  • fresh bleeding
  • fractures
  • hematomas
  • menstruation (vasodilation)

Transdermal application of drugs[edit | edit source]

The principle is the propagation of ultrasonic waves through a liquid, which is associated with the creation of pseudocavitation (tiny bubbles) that move chaotically and, after reaching a certain size, destabilize and collapse. At the same time, this creates a kind of jet microflow in the surrounding liquid, which causes abrasion of the upper layers of the skin. In addition to this abrasion, there is apparently also an increase in permeability of the phospholipid bilayer of biological membranes, which together results in a higher permeability of the skin even for high-molecular substances, which does not otherwise pass through her. Greater permeability of the skin is achieved by simultaneously acting on it with waves of two frequencies in the range from 20 kHz to 3 MHz. High frequency ultrasound creates additional bubbles that quickly burst in the presence of low frequency ultrasound waves. In addition, high-frequency ultrasound limits the movement of the bubbles to the sides and thus keeps them at the target location, which leads to a more uniform abrasion.

Such a liquid can be a liquid medicine applied to the skin. This is used (so far, rather experimentally) for chronic administration of drugs such as insulin, which previously had to be administered only by injection. The therapeutic effect thus lies in the minimal invasiveness of such an application. The abrasive effect of ultrasound is temporary and painless, the top layer of the skin (stratum corneum epidermis) regenerates quickly (within a few hours) from the basal layer of the epidermis.

Other uses[edit | edit source]

Furthermore, ultrasound is used to accelerate the healing of bone fractures, when joining bone fragments with a synthetic binder (the use of ultrasound shortens the setting time from days to seconds due to the local effect of higher temperature), removing atherosclerotic plaques in blood vessels, during the therapy of prostate tumors, procedures on the ovaries and in plastic surgery (removal of fatty tissue by extrusion).


Links[edit | edit source]

Related Articles[edit | edit source]

External links[edit | edit source]

References[edit | edit source]

  1. a b c PODĚBRADSKÝ, George – VAŘEKA, Ivan. Physical Therapy I. 1st edition edition. Grada Publishing, 1998. 
  2. ROSINA, Joseph. Medical Biophysics. 1. edition. Manus, 2000. ISBN 80-902318-5-3.
  3. a b c KUBÍNEK, Roman. University of Palacky Olomouc [online]. [cit. 2013-01-05]. <http://apfyz.upol.cz/ucebnice/details/ultrazvuk_terapie.pdf>.

References[edit | edit source]

  • PODĚBRADSKÝ, George – VAŘEKA, Ivan. Physical Therapy. Part 2. 1. edition. Grada, c1998. ISBN 80-7169-661-7.
  • ROSINA, Joseph. Medical Biophysics. 1. edition. Manus, 2000. 0 pp. ISBN 80-902318-5-3.